EasyManua.ls Logo

Keysight N5183B MXG - Parameter Errors (-220) - Spread Factors and Arm;Trigger Deadlocks; Spread Factor and Deadlock Parameter Errors

Keysight N5183B MXG
155 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight X-Series Signal Generators Error Messages 101
Error Messages
List of Error Messages
-220 Parameter error; Spread factor exceeds maximum allowable spread factor
The spread factor exceeds the maximum allowable spread factor. The
maximum allowable spread factor is currently 256.
-220 Parameter error; Spread factor not valid
The spread factor should be a power of 2 that is between 4 and 256.
-220 Parameter error; {name} The modulation type has been defaulted.
The specified file is not a modulation file type so the modulation
type has been defaulted.
-215 Arm deadlock
The arm source for the initiation of a measurement is set to
GET and a subsequent measurement query is received. The
measurement cannot begin until a GET is received, but the
GET would cause an INTERRUPTED error.
-214 Trigger deadlock
A trigger source for the initiation of a measurement is set to GET
and a subsequent measurement query is received. The measurement
cannot begin until a GET is received, but the GET would cause an
INTERRUPTED error.
-213 Init ignored
A request for a measurement initiation was ignored as another

Related product manuals