100 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
A program data element related error has occurred. This
particular error message is used if the device cannot detect
a more specific error described for errors -221 through -229.
-220 Parameter error; Channel Code exceeds spread factor.
The specified channelization code is too large. The channelization
code should always be at least one less than the spread factor.
-220 Parameter error; Data rate is not valid
The data rate is not contained in the defined data rate table.
-220 Parameter error; Real-time application has supplied a parameter
({name}) that is not supported for its type.
Real-time application has supplied a parameter that is not
supported for its type.
-220 Parameter error; Real-time application {name} has not supplied
parameter settings for {name}: default values are being used.
Real-time applications need to supply all parameter settings before
they can be turned on.
-220 Parameter error; Real-time application {name} is not allowed to send
legacy parameter settings for {name} when the instrument is not in
backwards compatibility mode.
Real-time applications should not be sending legacy parameter
settings when the instrument is not in backwards compatibility
mode.