Keysight X-Series Signal Generators Error Messages         103
Error Messages
List of Error Messages
            A command is not executable while the device is in local mode
            due to a hard local control (see IEEE 488.2, 5.6.1.5). For
            example, a device with a rotary switch receives a message
            which would change the switch's state, but the device is in
            local so the message cannot be executed.
-200 Execution Error
            For devices that cannot detect more specific errors, this
            code indicates only that an execution error as defined in
            IEEE 488.2, 11.5.1.1.5 has occurred.
-200 Execution error; cannot transmit paging message to Baseband generator.
            There is an error with the CDMA2000 or 1xEV-DV hardware. Try
            turning CDMA2000 or 1xEV-DV mode off and on again to reset the
            baseband generator card. If the problem persists, report this error 
            to service office.
-200 Execution error; cannot write user file to Baseband generator.
            There is a hardware error associated with the current CDMA format
   (cdma2000, WCDMA or 1xEV-DV). Try turning the current CDMA format
            mode off and on again to reset the baseband generator card. If the
            problem persists, report this error to service office.
-200 Execution error; OPC Query Timeout Error: output signal is unsettled.
     *OPC? timed out waiting for the signal to settle. This could be caused
     by an unlocked or unleveled condition.
            OPC query timeout occurred due to the signal being persistently
            unsettled. This happens when the instrument is unlocked or