110 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
More parameters were received than expected for the header. For
example, the *ESE common command only accepts one parameter, so
receiving *ESE 0,1 is not allowed.
-105 GET not allowed
A Group Execute Trigger was received within a program message
(see IEEE 488.2, 7.7). Correct the GPIB controller program so
that the GET does not occur within a line of GPIB program code.
-104 Data type error
The parser recognized a data element that is not allowed.
For example, numeric or string data was expected, but block
data was encountered.
-103 Invalid separator
The parser was expecting a separator and encountered an illegal
character. For example, the semicolon was omitted after a
program message unit.
-102 Syntax error
An unrecognized command or data type was encountered. For
example, a string was received when the device does not
accept strings.
-102 Syntax error; Bad compatibility language character {name}
In 8656/57 compatibility mode, illegal language input was received.