Keysight X-Series Signal Generators Error Messages         139
Error Messages
List of Error Messages
630 Arb Sequence Error; Single wave segment length must be at least 60
     samples. Value set to minimum.
            This error is reported when the user attempts to specify a single
            waveform sequence in which a segment is less than 60 samples. Due
            to hardware limitations of the sequencer, segments must be at least 
            60 samples long.
630 Arb Sequence Error; Single wave segment offset must be even. Value set
     to the next lower, even value.
            This error is reported when the user attempts to specify a single
            waveform sequence in which a segment begins at an odd sample
            offset. Due to hardware limitations of the sequencer, segments must 
            begin on an even sample pair boundary.
630 Arb Sequence Error; Unable to reset sequence on baseband board after
     change in file. {name}
            A change was made to a waveform file being referenced by the
            sequencer that has made the current sequence invalid. It is likely
            that the file contains too few samples. Correct the file size or
            the lengths and offsets in single waveform sequences and reenable
            the personality. {name} is the name of the file that caused the
            problem.
632 Power search failed. Power Unleveled.
            While executing power search, the level meter closed loop reference 
            value failed to settle. This event indicates that the power is in
            an unstable condition. The power will be set to the last properly
            leveled power.