Keysight X-Series Signal Generators Error Messages 139
Error Messages
List of Error Messages
630 Arb Sequence Error; Single wave segment length must be at least 60
samples. Value set to minimum.
This error is reported when the user attempts to specify a single
waveform sequence in which a segment is less than 60 samples. Due
to hardware limitations of the sequencer, segments must be at least
60 samples long.
630 Arb Sequence Error; Single wave segment offset must be even. Value set
to the next lower, even value.
This error is reported when the user attempts to specify a single
waveform sequence in which a segment begins at an odd sample
offset. Due to hardware limitations of the sequencer, segments must
begin on an even sample pair boundary.
630 Arb Sequence Error; Unable to reset sequence on baseband board after
change in file. {name}
A change was made to a waveform file being referenced by the
sequencer that has made the current sequence invalid. It is likely
that the file contains too few samples. Correct the file size or
the lengths and offsets in single waveform sequences and reenable
the personality. {name} is the name of the file that caused the
problem.
632 Power search failed. Power Unleveled.
While executing power search, the level meter closed loop reference
value failed to settle. This event indicates that the power is in
an unstable condition. The power will be set to the last properly
leveled power.