EasyManua.ls Logo

Keysight N5183B MXG - Power Search Errors (633 to 637); Power Search Failure Conditions

Keysight N5183B MXG
155 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
140 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
633 Power search failed. Reference reading unsettled.
While executing power search, the level meter closed loop reference
value failed to settle. This event indicates that the power is in
an unstable condition. The power will be set to the last properly
leveled power.
634 Power search failed. ALC open loop unsettled.
While executing power search, the level meter open loop voltage
failed to settle. This event indicates that the power is in an
unstable condition. The power will be set to the last properly
leveled power.
635 Power search failed. ALC abus voltage unsettled.
While executing power search, the level meter ALC voltage failed to
settle. This event indicates that the power is in an unstable
condition. The power will be set to the last properly leveled
power.
636 Power search failed. DAC value exceeded range.
While executing power search, the value requested for the offset
compensation DAC was beyond its available range. This event
indicates a low power level that cannot be compensated with the
hardware. The power will be set to the last properly leveled power.
637 Power search failed. Excessive iterations.
While executing power search, the number of iterations in the power

Related product manuals