142 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
in secure mode can only be increased.
644 Firmware extension module error; Cannot load extension module due to
missing file ({name}).
This error indicates that specified file containing code for self
test or calibrations is missing. Report this error to service
office.
644 Firmware extension module error; Failed access to operation in
calibration extension module({value}).
This error indicates that there was a problem accessing a specified
operation in the calibration extension module. This can occur if
the calibration extension module is corrupt or missing or does not
contain the referenced operation.
644 Firmware extension module error; Specified operation number ({value})
does not exist.
There was a problem accessing a specific operation in one of the
extension modules. This occurs if the extension module is corrupt
or missing or does not contain the referenced operation.
645 License specification error; {name}
The file may be corrupt or the firmware may need to be upgraded to
understand the license specification.
646 Default path not found on USB; To use the external media as the user
file system, you must choose an existing path on the external media.
Current path is '{name}'.