18 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
-315 Configuration memory lost; Persistent state preset. Using factory
defaults.
The persistent state has been forced to return to factory preset
values.
-313 Calibration memory lost; Cal not found and was created: {name}
The specified cal file was not found so it was created by the
system.
-313 Calibration memory lost; Cal out of mem: {name}
Inadequate memory for specified cal file.
-313 Calibration memory lost; Cal resizing: {name}
Inadequate memory to resize specified cal file.
-313 Calibration memory lost; Cal restored from backup: {name}
A specified cal file was restored from backup.
-313 Calibration memory lost; Cannot restore cal from backup:
{name}err={value}
Cannot restore specified cal from backup because of a file error.
-313 Calibration memory lost; Cannot store cal to backup: {name}err={value}
Cannot store specified cal to backup because of a file error.