Keysight X-Series Signal Generators Error Messages 17
Error Messages
List of Error Messages
fragmentation). Report this error to service office. The instrument
is still functional.
-321 Out of memory; Unable to display timeslot window.
The instrument was unable to create part of the graphical user
interface due to an inability to allocate memory (possibly due to
fragmentation). Report this error to service office. The instrument
is still functional.
-321 Out of memory; User flatness subsystem unable to operate.
There is insufficient working memory available. Report this error
to service office.
-321 Out of memory; {name}
The header file to be written is too big to fit on the media. The
new information was not written and the original header is
untouched.
-321 Out of memory; {name} display cannot function.
There is insufficient working memory available. Report this error
to service office.
-315 Configuration memory lost; Persistent state checksum is bad. Using
factory defaults.
The persistent state is corrupt and had to be reinitialized with
the factory preset values.