Loading...
46 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
-224 Illegal parameter value; The specified option is not degradable.
An error occurred when the user attempted to degrade an option that
is not able to be degraded.
-224 Illegal parameter value; The specified option was not found.
An error occurred when the user attempted to degrade an option that
doesn't exist.
-224 Illegal parameter value; Total user file bits must be greater than 0.
An attempt was made to select a user file that contains 0 bits.
Select another user file that contains at least 1 bit.
-223 Too much data
A legal program data element of block, expression or string type
was received that contained more data than the device could handle
due to memory or related device-specific requirements.
-223 Too much data; The current power setting results in too many data
points. A new symbol map cannot be generated. The current channel
power has been reverted to its previous value.
Due to hardware limitations, only a certain number of unique data
points can be created for the IQ map. This channel power level has
resulted in an IQ map that exceeds the hardware limit when turned
on, and cannot be accommodated. The channel power has been
reverted to its previous value to retain a valid signal. To turn this
channel on, try a different power setting or turn some other
channel off.