Keysight X-Series Signal Generators Error Messages 55
Error Messages
List of Error Messages
-222 Data out of range; {name} clipped to upper limit.
An input value is above the maximum value allowed. In some cases
the maximum value is affected by other item setting due to
coupling.
-221 Settings conflict; (delay + width) must be less than the period and
less than (42s - width). For the second pulse in an Adjustable
Doublet, the delay must be less than (42s - (first_pulse_delay) -
(second_pulse_width)).
-221 Settings conflict; 3GPP Uplink AWGN refers PRACH Message Part, but
the
Message Part is set to off.
3GPP Uplink AWGN refers PRACH Message Part as E ref, but the
Message Part is set to off. To avoid this, set the Message Part to
be on.
-221 Settings conflict; 3GPP Uplink PRACH MessageTotalPower can be set
only
when Total is selected as PowerSetUpMode. Pp-m can be set only when
Pp-m is selected.
3GPP Uplink PRACH MessageTotalPower can be set only when Total is
selected as Power SetUp Mode. Pp-m can be set only when Pp-m is
selected as Power SetUp Mode.
-221 Settings conflict; 8757D system mode does not support the following
command: {name}
The instrument is connected to an 8757 network analyzer in smart
mode and cannot set this feature.