56 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
-221 Settings conflict; A change in the clock rate has caused the skew to
be outside the valid range. The skew has been set to zero.
When the frequency is changed and the skew setting is to high or
too low for the new frequency, this error is generated to report
that the skew has been re-set to zero.
-221 Settings conflict; A change to the modulation type has limited the
maximum symbol rate.
A change to the modulation type has limited the maximum symbol
rate.
-221 Settings conflict; All channels are off. Could not perform operation.
This error is reported when an attempt is made to balance powers
when all channels are off.
-221 Settings conflict; All timeslots with the payload (E) fields set to
EXT have been reset to PN9. EXT is incompatible with combined EDGE/GSM
mode.
EXT is incompatible with combined EDGE/GSM mode.
-221 Settings conflict; AM cannot be in exponential mode in the lowest
frequency band. AM has been switched to linear mode.
The lowest frequency band (9 kHz up to 5 MHz) does not have
exponential AM.
-221 Settings conflict; An external reference frequency other than 10MHz