Keysight X-Series Signal Generators Error Messages                   59
Error Messages
List of Error Messages
-221 Settings conflict; BBG data clock has been set to internal. Ext data
     bit clock does not work with data rates over 50 Mbps. Ext symbol clock
     is okay.
            Parallel mode does not work with an external bit clock. However it
            can be used with an external symbol clock.
-221 Settings conflict; BBG trigger source and/or type settings for {name}
     conflicted with Multi-BBG Sync and have been adjusted accordingly.
            BBG trigger types Continuous-Free Run and Gated are not available
            with Multi-BBG Sync. With Multi-BBG Sync type Slave, the only BBG
            trigger source permitted is EXTERNAL using PAT TRIG-1.
-221 Settings conflict; BCH Conflicts with another multiframe type.
            *** No documentation ***
-221 Settings conflict; BCH is allowed only in TimeSlot 0.
            *** No documentation ***
-221 Settings conflict; The bit rate has been set to 50 Mbps. Ext (serial)
     data in does not work with data rates above 50 Mbps.
            Ext (serial) does not work at rates above 50 Mbps. Ext parallel can 
            be used at higher bit rates.
-221 Settings conflict; Block size has been adjusted to match BlockSetSize
     / NumberOfBlks. But block size does not match the BlockSetSize
     exactly. Try another value.