68                               Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
            only one modulation at a time. If a modulation requests a source
            while it is in use by an active modulation, the first modulation is 
            turned off, and the second modulation is turned on.
-221 Settings conflict; External data cannot be selected for multiple
     channels. PN9 is automatically selected for {name}.
            The current hardware only supports one external data source at any
            given time. The <CHANNEL NAME> data source has been changed 
            PN9, the default data source.
-221 Settings conflict; The external input attenuator cannot compensate for
     the low signal voltage on the analog I/Q inputs.
            As the voltage level on the external I/Q inputs is decreased, the
            external input attenuation also decreases to keep the I/Q modulator 
            voltage level constant. When the attenuator reaches zero, it can no 
            longer provide compensation for the input voltage and further
            decreases in input voltage will result in decreased power out of
            the I/Q modulator. If the ALC is off this will be observable in the 
            RF output power. If the ALC is on, the ALC will compensate for the
            decreased input voltage and there will be no observable change in
            RF output power. For very low voltages the ALC will not be able to
            compensate and the RF output power will become unleveled.
-221 Settings conflict; FFCH/OCNS and QPCH cannot be on at the same time.
     QPCH has been turned off.
            QPCH is generated using both channel 3 and 4. Channel 3 has been
            turned on to generate FFCH or OCNS. Channel 4 can no longer
            generate QPCH. Therefore, QPCH must be turned off.