Keysight X-Series Signal Generators Error Messages 69
Error Messages
List of Error Messages
-221 Settings conflict; The Flat Zone Amplitude(Lmid) is coupled to the
Stop Frequency(f2). The Flat Zone Amplitude has been limited
appropriately for the new Stop Frequency.
-221 Settings conflict; FM and PM not allowed.
There is a hardware conflict between FM and PM. The most recently
requested modulation will be turned on. The previous modulation
will be turned off.
-221 Settings conflict; FM1/PM1 value set less than FM2/PM2 value.
FM2/PM2
changed to match FM1/PM1 value.
The deviation of FM2/PM2 must always be less than or equal to the
deviation settings for FM1/PM1. This error will be reported to the
queue when FM2/PM2 is enabled and FM1/PM1 is also enabled and an
adjustment to either FM1/PM1 deviation causes the FM1 or PM1
deviation to be less than the FM2 or PM2 deviation. It will also be
reported when FM1/PM1 is being turned on, and the last FM2/PM2
deviation setting is greater than the current FM1/PM1 deviation
setting. In both cases the FM2/PM2 deviation will be adjusted to
match the FM1/PM1 deviation.
-221 Settings conflict; FM2/PM2 value set greater than FM1/PM1 value.
FM1/PM1 changed to match FM2/PM2 value.
FM2/PM2 deviation must be FM1/PM1 deviation. This error is reported
when both FM1/PM1 and FM2/PM2 are enabled and an adjustment to
either FM2/PM2 deviation causes the FM2 or PM2 deviation to be
greater than the FM1 or PM1 deviation. It is also be reported as
FM2/PM2 is turned on, and the last FM1/PM1 deviation setting is