70 Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
less than the current FM2/PM2 deviation setting. In both cases the
FM1/PM1 deviation is adjusted to match the FM2/PM2 deviation.
-221 Settings conflict; FM/PM deviation decreased to account for FM rate
and ramp sweep settings.
The deviation and rate settings for FM and PM are interrelated when
ramp sweep is active. The deviation setting will be automatically
set to an acceptable value when the rate moves outside an
acceptable range.
-221 Settings conflict; FM/PM rate increased to account for deviation and
ramp sweep settings.
The deviation and rate settings for FM and PM are interrelated when
ramp sweep is active. The rate setting will be automatically set to
an acceptable value when the deviation moves outside an acceptable
range.
-221 Settings conflict; For the second pulse in an Adjustable Doublet, the
pulse width must be less than (42s - (first_pulse_delay)) and the
delay must be less than (42s - (first_pulse_delay) -
(second_pulse_width)).
-221 Settings conflict; Frame repeat changed to continuous.
Single is not a valid repeat setting for an external data source.
-221 Settings conflict; The frequency spacing has been limited by the
number of carriers.
The frequency spacing and the number of carriers must be such that