Keysight X-Series Signal Generators Error Messages 71
Error Messages
List of Error Messages
the bandwidth limitation of the baseband generator is not exceeded.
-221 Settings conflict; The frequency spacing has been limited by the
number of tones.
The frequency spacing and the number of tones must be such that the
80 MHz of bandwidth is not exceeded. If FS_MAX is the maximum
frequency spacing in MHz and N is the number of tones, then {FS_MAX
<= (80) / (N - 1)}.
-221 Settings conflict; The frequency spacing has been limited by the tone
alignment.
The frequency spacing and alignment must be such that the 80 MHz of
bandwidth is not exceeded and that no tone is more than 40 MHz from
the carrier.
-221 Settings conflict; FSK frequency deviation has been limited by the
sample rate.
The Sample Rate must be at least 16 times the Digital Mod FSK
frequency deviation. If this cannot be achieved without exceeding
the maximum sample rate for the current settings, the FSK frequency
deviation will be limited.
-221 Settings conflict; FSK frequency deviation has been limited by the
symbol rate.
The maximum FSK frequency deviation depends on the symbol rate.
-221 Settings conflict; If amplitude reference state is on, power units
must be dB. If amplitude reference state is off, allowable power units