Keysight X-Series Signal Generators Error Messages 83
Error Messages
List of Error Messages
Power Meter. Leveling set to internal.
The Power Meter Servo mode could not be enabled. ALC Leveling Mode
set to internal. Please check the Power Meter.
-221 Settings conflict; The power search reference is not valid with the
selected I/Q source. The power search reference has been adjusted
accordingly.
The power search reference types are only valid when I/Q is on. If
the I/Q source is external, the rms power search reference is not
available.
-221 Settings conflict; Pre-FIR samples are not allowed when DSIM is
outputting AWGN or Phase Noise modulation modes. Pre-FIR samples are
also not allowed with constant envelope modulation in Real-Time
Custom.
Pre-FIR samples are not allowed when DSIM output (option 003) is in
operation and the modulation mode that it is outputting doesn't
support pre-FIR samples. Pre-FIR samples are also not allowed with
constant envelope modulation in Real-Time Custom.
-221 Settings conflict; Pulse Modulation is not compatible with Marker
Beacon Pulsed mode. Pulse Modulation has been turned off.
The Marker Beacon Pulsed mode adjustes the pulse modulator;
therefore, Pulse Modulation and Marker Beacon Pulsed mode ON are
not compatible.
-221 Settings conflict; Pulse period set less than pulse width value. Pulse
width changed to match period value.