Keysight X-Series Signal Generators Error Messages 85
Error Messages
List of Error Messages
-221 Settings conflict; RACH is only defined for Radio Configuration 1 and
2. You must set one of these Radio Configurations before configuring
or using this code channel.
An attempt was made to setup the R-ACH while in a radio
configuration that does not support this channel. This channel is
only available in RC1 and RC2. If the R-ACH is the desired channel,
you must change to the proper radio configuration before
continuing.
-221 Settings conflict; Ramp Sweep and FM/PM cannot be on at the same
time.
The modulation type FM and Ramp Sweep Mode cannot be used at the
same time. This error is reported when the user attempts to
activate fm modulation while the sweep is activated. The mode that
was most recently requested will be turned on and the mode that was
on at the time will be turned off.
-221 Settings conflict; Ramp Sweep cannot be on during DCFM/DCPM Cal.
A DCFM/DCPM Cal cannot be done in Ramp Sweep Mode. This error is
reported when the user attempts to perform a DCFM/DCPM calibration
while sweeping.
-221 Settings conflict; Ramping step should be 0 when Number of Preamble is
Infinity.
When the repetition number of PRACH Preambles is set to "Infinity",
the ramping step power of Preambles must be "0" dB. This is
equivalent to constant Preamble power configuration.