Keysight X-Series Signal Generators Error Messages                   87
Error Messages
List of Error Messages
     Multiplexed operating mode for the AUX port: the mode has been changed
     to Dedicated.
            The real-time application does not support Multiplexed operating
            mode for the AUX port.
-221 Settings conflict; Real-Time AWGN in ARB formats cannot be used when
     the RMS value of the waveform is unknown. Please see any previous
     error message for why the RMS could not be calculated. The noise has
     been turned off.
            Real-Time AWGN in ARB formats cannot be used when the RMS value
            of the waveform is unknown. Please see any previous error message for
            why the RMS could not be calculated. The noise has been turned off.
-221 Settings conflict; Real-Time mode {name} has turned on unexpectedly.
            The Real-time data transform mode has turned on unexpectedly.
-221 Settings conflict; Real-Time {name} mode could not be turned on
     because there is no format currently on. Please turn on a format
     before turning on {name}.
      When turning on a Real-Time data transform mode, a check is made to 
            see if a format is currently on. This error is generated when no
            format is found.
-221 Settings conflict; Real-Time {name} mode was forced off because the
     current format is not supported.
            Real-Time data transform mode had to be turned off as the current
            format is not supported with it.