Keysight N5183B MXG - Settings Conflict Errors (-221) - Turbo Coding and Puncture Rate; Turbo Coding and Puncture Rate Conflicts

Keysight N5183B MXG
155 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight X-Series Signal Generators Error Messages 95
Error Messages
List of Error Messages
while having any internal signals routed to it.
When Trig1 or Trig2 BNC is routed to the Sweep or Point Trigger no
other internal signals can be routed to Trig1 or Trig2. You must
first remove the routing to the Sweep or Point Trigger before
routing internal signals to Trig1 or Trig2 BNCs.
-221 Settings conflict; Trigger Source setting now conflicts with Dynamic
Switching. Segment Order has been changed to LINEAR to accomodate
request.
Trigger Source setting now conflicts with Dynamic Switching.
Segment Order has been changed to LINEAR to accomodate request.
-221 Settings conflict; Turbo Coding is not available in the current
configuration. Please review your manual for valid configurations.
Turbo Coding is not available in the current
configuration. Please review your manual for
valid configurations.
-221 Settings conflict; Turn on channel #{value} caused the puncture rate
to exceed the maximum specified rate for at least one transport
channel. Transport channel #{value} has been turned off.
The puncture rate of each TrCH should be smaller than 60%, which is
displayed on UI screen as "Max Puncture" value. Max Puncture is
calculated as (1 - Puncturing Limit) * 100. Puncturing Limit is
defined at 3GPP STD TS 125 331. In this case, data bits number
before rate matching is increased because of TrCH addition, and
symbol rate of physical channel is not changed. So puncture rate is
increased. To avoid the error, increase the symbol rate of physical

Related product manuals