94                               Keysight X-Series Signal Generators Error Messages
Error Messages
List of Error Messages
            or 0. TGPL1 * 15 >= TGSN + TGL1
-221 Settings conflict; TGD parameter is not set to Undefined or 0. TGPL2 *
     15 >= TGSN + TGL2 + TGD TGD parameter is set to Undefined or 0. TGPL2
     * 15 >= TGSN + TGL1
            TGPL2 is the duration of transmission gap pattern 2. The gap
            pattern consists of TGSN, TGD, TGL1 and TGL2. The gap pattern can't 
            be generated if the TGPL2 value is too short. TGPL2 should meet the 
            condition as below; TGD parameter is not set to Undefined or 0.
            TGPL2 * 15 >= TGSN + TGL2 + TGD TGD parameter is set to Undefined
            or 0. TGPL2 * 15 >= TGSN + TGL1
-221 Settings conflict; the Transport Position must be FIXED when
     compressed mode is on.
            The Transport Position must be FIXED when compressed mode is on.
-221 Settings conflict; Transmit Diversity should be NONE when Compressed
     Mode is on or OutOfSyncTest is on.
            Transmit Diversity should be NONE when Compressed Mode is on or
            OutOfSyncTest is on.
-221 Settings conflict; The TrCH must be selected in Data on DPDCH or Msg
     Data physical channel setup.
            The TrCH isn't selected in DPDCH(Msg Data) physical channel setup
            in spite of 3GPP Uplink AWGN E bRef(E ref) refers to DCH(RACH). Set 
            TrCH as the data type.
-221 Settings conflict; Trig BNC cannot be routed to Sweep or Point Trigger