Option Esc External Source Control
Using Option ESC with the Keysight MXG Signal Sources.
Lan
The X-Series with Option ESC controls the MXG through a LAN connection based 
on TCP/IP protocol. The TCP/IP protocol can only be established with correct IP 
addressing. 
NOTE You must be logged in as an Administrator to change the TCP/IP properties. This 
requires the use of a USB mouse and keyboard.
X-Series Instrument
NOTE Assure only ONE correction file is turned on because it is possible to turn on 
multiple correction files, and if some of the files share the same frequency points, the 
correction that results for those shared frequency points will cause measurement 
errors. Therefore assure that only the correction file required for the measurement is 
turned on, and turn off all other corrections. 
Loading conversion loss data for the PXA Signal Analyzer
The conversion loss data supplied with your mixer can be loaded into your signal 
analyzer from one of two sources:
8 Turn on the Signal ID 
function
• Press Input Output, External 
Mixing, Signal ID Mode, Image 
Suppress, Signal ID On. 
This enables you to identify true 
signals from images and 
harmonics.
See “Signal ID” on page 182 
for more information.
Step Action Notes
Step Action
1 Log off as default user.  •Click Start, Log Off, Log Off
2 Log on as an administrator. • At the log in prompt enter: 
User name: administrator 
Password: agilent4u
3 Assign an IP adress and a subnet mask.