Making Distortion Measurements
Identifying Analyzer Generated Distortion
Figure 6-1  Harmonic Distortion
5 Change the center 
frequency to the value 
of the second harmonic.
• Press Peak Search, Next Peak, 
Mkr→CF.
6 Change the span to 
50 MHz and re-center 
the signal.
a. Press SPAN X Scale, Span, 50, 
MHz.
b. Press Peak Search, Mkr→CF.
7 Set the attenuation to 
0 dB.
• Press AMPTD Y Scale, 
Attenuation, 0, dB.
8 Save the trace data in 
trace 2.
• Press Trace/Detector, Select 
Trace, Trace 2, Clear Write.
9 Allow trace 2 to update. • Press Trace/Detector, 
View/Blank, View, Trace On.
Minimum of two sweeps.
10Place a delta marker on 
the harmonic of trace 2.
• Press Peak Search, Marker Delta. The analyzer display shows the 
stored data in trace 2 and the 
measured data in trace 1. The 
ΔMkr1 amplitude reading is the 
difference in amplitude between 
the reference and active markers.
Step Action Notes