6 Installation Note N9030-90045
Option BBA, Analog Baseband I/Q Inputs Retrofit Kit
Tools Required
PXA Service Guide as N9030A-0BW or N9030-90030
T-10 Torx Driver
T-20 Torx Driver
5/16-inch torque wrench
1/4-inch open end wrench
1/4-inch socket on 4 lb torque wrench
scissors or knife
diagonal cutters
USB storage device with > 2 GB free memory
WARNING Before you disassemble the instrument, turn the power switch to Standby,
wait until the instrument has completely shut down, and unplug the
instrument. Failure to unplug the instrument can result in personal injury.
CAUTION Electrostatic discharge (ESD) can damage or destroy electronic components. All
work on electronic assemblies should be performed at a static-safe workstation.
Refer to the documentation that pertains to your instrument for information about
static-safe workstations and ordering static-safe accessories.
Agilent Service Center or Customer Calibration Lab:
The following kit items are not installed in the analyzer upgrade and need to be shipped back to the
end user for their use:
USB storage device
5063-2143, Probe tip BNC adapter
8120-2682, cable
Entitlement Certificate

Other manuals for Keysight N9030A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9030A and is the answer not in the manual?

Keysight N9030A Specifications

General IconGeneral
BrandKeysight
ModelN9030A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals