Noise Figure Mode (Option 356 – CPU2 Only)
How to use Auto-Integration
— ENR Table Edit/Save/Recall 
— Edit ENR Table
 
— Enter the serial number and model number of your noise source. If 
desired, enter other information (e.g., Options, Cal Date, etcetera).
Enter a value using the numeric keypad, the   arrows, or the 
rotary knob.
— Done
 to finish table and return to previous menu.
3. Save the ENR table: 
If you are not familiar with saving and recalling files, refer to Chapter 28, 
“File Management.” and then to “Saving and Recalling Files” on page 544.
Choose:
— Save ENR Table
 
Enter the filename of your noise source.
Enter a value using the numeric keypad, the   arrows, or the 
rotary knob.
—
Done to finish table and return to previous menu.
4. Enter the noise figure measurement’s frequency, number of points, and 
span settings: 
— Freq/Dist
 > Center > 50 GHz
— Sweep 3
 > More > 401 points 
— BW 2 > Span > 0 Hz 
5. Define the noise figure’s integration settings:
Meas Setup 4
 > Integration
Then choose
—  Mode Auto
 (default)
— Jitter Goal
 > 0.10 dB 
— Max. Time/Pt
 > 300s 
— Jitter Warning ON
 (default) 
— Back
 to return to previous menu
Press Uncertainty
 then
— Uncert. Bars ON
6. Leave the setup to run for desired time.