Keysight N9927-90001 User’s Guide                91
NA (Network Analyzer) Mode
NA Mode Settings
Multi-Trace Configurations
In NA Mode you can display multiple traces on the FieldFox screen. 
The graphic above shows a x3H configuration. Tr2 is the ACTIVE trace as 
indicated by the highlighted Tr2.
Trace Setting Notes
— The Frequency Range, IF BW, Resolution, Average, and Output Power 
settings are common for all displayed traces.
— All other trace settings, such as measurement, format, and limit lines, are 
applied individually to the ACTIVE trace in the same manner as when a 
single trace is present.
— By default, a marker is created on ALL traces. However, they can be created 
individually by disabling Coupled Markers. Learn more in 
“Coupled Markers 
(NA Mode)” on page 487
.
IMPORTANT! For CAT and NA modes, limit lines do not apply  
where F1 = F2.