Conducted Emissions Measurement Examples
Making Conducted Emission Measurements
7 Change EMI Average 
detector to compare to Limit 
Line 2
a. Press Meas Setup, 
Detectors (Measure).
b. Select Detector, Detector 
3.
c. Press Limit for Δ, Limit 2, 
Enter.
 
 
 
 
 
A dialog box may appear, 
Changing limit for delta 
will discard delta 
values, are you sure you 
want to do this? Please 
press Enter of OK to 
proceed. Press ESC or 
Cancel to close this 
dialog.
8 Load correction factors for 
the LISN
a. Press Recall, Data, 
Amplitude Correction, 
Select Correction, 
Correction 1.
b. Press Preloaded Limits.
c. Press Return, Open.
d. Select LISN-10A 
(9 kHz to 
30MHz).csv, 
Open.
This places the corrections for the 
LISN in Amplitude Correction 1.
 
 
These correction factors 
compensate for the losses of the 
LISN.
9 Insure that the correction 
factors are on
a. Press Input/Output, More 
1 of 2, Corrections, 
Correction 1, On.
10Update the scan a. Press Sweep/Control, 
Start.
View the ambient emissions (with 
the DUT off). If emissions above the 
limit are noted, the power cord 
between the LISN and the DUT may 
be acting as an antenna. Shorten 
the power cord to reduce the 
response to ambient signals.