Disturbance Analyzer (Click) Measurements
Setup Table Menu Features
Factor f
For certain types of products that must be cycled to emit discontinuous disturbances
(rather than run continuously), CISPR 14-1 requires users to operate the product over
enough cycles to product 40 clicks.
Factor f is used to calculate the click rate for these types of devices. See CISPR 14-1,
Annex A Table A.2 for the factor to use for your specific DUT.
NOTE This information is given as an example. CISPR 14-1 is the reference document for
disturbance measurement requirements. Refer to CISPR 14-1 to identify the test
requirements for your specific DUT.