Installation Note N9030-90036 9
Option CR3, or CRP, or ALV Installation Kit
Option ALV, Fast Log Video
1. Preset X-Series analyzer. Assure mode is Spectrum Analyzer.
2. Enable Option ALV. Press
Input/Output, More, Output Configuration, Aux IF Out, Fast Log
Video
.
3. Connect a signal source configured for a pulse output to the X-Series analyzer RF input port. The
frequency of the pulse should be above 200 MHz. Example settings are:
Pulse frequency: 400 MHz
Amplitude: 20 dBm
Pulse period: 80 μsec
Pulse width: 40 μsec
4. Spectrum analyzer settings:
Center frequency: 400 MHz
Span: 0 Hz
Bandwidth: 3 MHz
5. Connect an oscilloscope to the spectrum analyzer rear panel Aux IF Out port.
Oscilloscope settings:
500 mV/div
10 μsec/div
Adjust oscilloscope to view the pulse on the oscilloscope. Pulse should measure approximately
1.3 VP-P