5
Other cautions
When you use this equipment, be careful of the following things.
This equipment is Lumitester Smart/PD-30/PD-20 exclusive use. Do not insert a
light source or blank tube in Lumitester PD-
10. Or do not insert to other company
products.
Do not move this equipment during measurement. It may
cause an error of
measurement.
Do not incline this
equipment during measurement. It may cause a measured
value error.
Be sure to take out a light source after measurement.
Keep this equipment at least 50cm away from the electromagnetic noise source.
Do not expose this equipment to liquids, reagents, and organic solvents. That may
cause failure, discoloration and modification. If they are attached, wipe off immediately.
Use it in the range of temperature (10-30
ºC) and humidity (20-85%).
*In case of Lumitester Smart/PD-
30, turn off the temperature compensation
setting.
Do not use it at the place where temperature changes violently.
Do not use it at a place with sunlight or strong light. It may cause
an error of
measurement.
Do not expose it to steam. Use it at the place without dewing.
Do not expose it to corrosive gas.
Use it at a place without vibration.
Use it at a place with little dust.
Do not put on unstable place or a dangerous place.
Do not fall it or give a strong shock.
Avoid the place where the wind of an air-conditioner hits directly.