K110sm7e1.doc
7-7
7. 3. 4 Some image is lost at the boundary of Image
Blocks
Cause Checking
order
Checking Result Treatment
Calibration 1 Can you fix the problem if you make
Position?
(Refer to [8.4.4 Position].)
Yes OK
7. 3. 5 Vertical image gap between Image Blocks
Cause Checking
order
Checking Result Treatment
Calibration 1 Can you fix the problem if you make
Position?
(Refer to [8.4.4 Position].)
Yes OK
7. 3. 6 Image quality is not good
Cause Checking
order
Checking Result Treatment
Scan Glass 1 Is there any dirt or damage on the Scan
Glass?
Yes Clean or replace it.
Resolution 2 Is the resolution setting proper? No Adjust it properly.
7. 3. 7 Uneven density
Cause Checking
order
Checking Result Treatment
Calibration 1 Can you fix the problem if you make
Shading and Color Correction?
(Refer to [8.4.1 Shading] and [8.4.2
Color Correction].)
Yes OK