KL1 Operating Manual
RETAIN THESE INSTRUCTIONS FOR REFERENCE
www.capsahealthcare.com

Other manuals for Kirby Lester KL1

Questions and Answers

  • M
    Marcus ScottAug 3, 2025
    Why does my Kirby Lester Medical Equipment count by itself?
    • S
      Samantha CopelandAug 3, 2025
      If your Kirby Lester Medical Equipment is counting by itself or displaying errors like E-1, E-9, or O-S, it's likely due to dust or residue on the glass detector windows, or the windows are cracked. Try cleaning the glass detector windows thoroughly. If the glass is cracked, the unit needs to be sent in for repair.
  • S
    Samantha RobertsonAug 7, 2025
    Why is my Kirby Lester Medical Equipment overcounting pills?
    • M
      Mark CarpenterAug 7, 2025
      If your Kirby Lester Medical Equipment overcounts, meaning it counts more pills than you poured, the cause might be dust or residue on the glass detector windows, cracked windows, or pill fragments that were poured through. You should thoroughly clean the glass detector windows and remove any pill fragments. If the glass is cracked, send the unit in for repair.
  • L
    Lisa WilsonAug 11, 2025
    What to do if my Kirby Lester Medical Equipment undercounts pills?
    • C
      Crystal StephensonAug 11, 2025
      If your Kirby Lester Medical Equipment undercounts, meaning it counts fewer pills than you poured, it could be because you are pouring too fast or directly into the funnel hole. Try slowing down your pouring speed, so the 'Pour Slower' indicator does not activate.
  • S
    Sarah MillerAug 16, 2025
    Why is my Kirby Lester KL1 Medical Equipment beeping?
    • S
      Sydney KellyAug 16, 2025
      If your Kirby Lester KL1 Medical Equipment is beeping, it means medications are still in the tray. Simply empty the tray.
  • D
    Debra RiveraAug 20, 2025
    What to do if my Kirby Lester Medical Equipment display flashes '----'?
    • G
      Gregory TrujilloAug 20, 2025
      If the display on your Kirby Lester Medical Equipment flashes "----" instead of "0000" when the tray is pushed in, it could be that the tray isn't fully inserted or the tray sensor window is dirty. Make sure the tray is fully inserted and clean the tray sensor window.
  • K
    Kimberly RobertsonAug 25, 2025
    Why is the display flashing on my Kirby Lester Medical Equipment?
    • L
      lopezrobertAug 25, 2025
      If the display on your Kirby Lester Medical Equipment is flashing, it indicates that the tray is missing. Empty the tray, then fully insert it.
  • K
    kevin57Aug 28, 2025
    Why does my Kirby Lester Medical Equipment read '0000' with the tray not fully inserted?
    • D
      Daniel BuchananAug 29, 2025
      If your Kirby Lester Medical Equipment displays "0000" even when the counting tray is not fully inserted, the tray sensor might be dirty or damaged. First, try cleaning the tray sensor window. If that doesn't work, the unit may need to be sent in for repair if the tray sensor is damaged.

Summary

Installation Guide

Counting Procedures

Pour Slower Message

Alerts user when pouring speed is too fast, requiring slowdown for accurate counting.

Partial Pill Message

Alerts user to detect partial tablets/fragments, prompting inspection and removal.

Special Mode

Mode for counting unique, odd-shaped, or semi-translucent tablets and capsules.

Inventory Mode

Mode for counting large quantities of tablets/capsules exceeding tray capacity.

Cleaning and Maintenance

Funnel Cleaning

Instructions for cleaning the funnel and channel using mild soapy water.

Glass Detector Windows Maintenance

Guide to cleaning glass detector windows to maintain counting accuracy.

Counting Tray Cleaning

Procedure for cleaning the counting tray with mild soapy water.

Tray Sensor Cleaning

Instructions for cleaning the tray sensor window to ensure proper seating detection.

Technical Specifications

Safety Precautions

Troubleshooting Common Issues

Kirby Lester Service Information

Kirby Lester KL1 Specifications

General IconGeneral
BrandKirby Lester
ModelKL1
CategoryLaboratory Equipment
LanguageEnglish