K394XX Water Separability Tester
Operation and Instruction Manual
K394XX-Manual -12-
Figure 11 . Stirrer Not Engaged
Figure 12. Stirrer Engaged
Figure 13: Run Screen
5. After pressing the START TEST button, the
Running Screen (Figure 13) will appear.
6. To pause the test, press the PAUSE button.
Pressing the PAUSE button twice will end
the test and lead the user back to the set up
screen. Press the CONTINUE button to
continue running the test.
6 Safety Features
The Koehler Water Separability Apparatus is
equipped with several safety and protection
features, which are described in the following
sections.
WARNING: The bath is NOT Explosion
Proof.
6.1 Over Temperature Protection
The Koehler Water Separability Apparatus is
equipped with Over-temperature Protection
(OTP) circuitry, which prevents the unit from
exceeding unsafe operating temperatures. If the
unit cannot maintain the set point temperature
and begins to decline, the OTP circuitry may
have been activated. Please follow these steps.
1. Turn off the unit power switch and disconnect
the line cord.
2. Determine the source of the problem and
correct the situation.
3. Restart the unit. Monitor the operations to
ensure that the unit is operating properly. If
you are still experiencing trouble, please
contact Koehler technical service for
assistance.
6.2 Over Power Protection
The Koehler Water Separability Apparatus is
equipped with Over-power Protection circuitry,
which prevents the unit from unsafe electrical
conditions. If power to the unit is lost, then turn
off the main power and turn it back on again.
The main power switch also functions as a
circuit breaker.