MJ400080F1
7. Other settings
Press both SET and
for about 2 seconds
Select and set GAL or L
by pressing or
Press SET
Set the type by pressing
or
Press SET
Set ammeter output value
when analog output is zero
by pressing or
Press SET
Connect ammeter
(or corresponding
resistance/voltmeter)
to analog output line
Set ammeter output value
when analog output FS
by pressing or
Press SET
Set to actual flow device
and supply 100% FS
corresponding flow
Set ammeter output value
when flow 100% FS
by pressing or
Press SET
Flow unit setting mode
Measuring mode
Flow unit setting completed
Type setting mode
Type setting completed
Output zero setting mode
Output zero setting completed
Output span setting mode
Output span setting ended
Flow span setting mode
Flow span setting ended
Write in EEPROM