aeroplus E
Maintenance
39
Component
Hospital use
action
Out-of-hospital use
action
FireSafe check valve
Replace every 6 months
Oxygen mask
Weakly cleaning and
disinfection
Replace monthly
Weakly cleaning
Replace monthly
Device inlet filter
Replace yearly / after 5000 operating hours
9.3.2 With changing patients
Component
Required action
aeroplus E
Cleaning and disinfection
Humidifier
Dispose of
Holder for humidifier
Cleaning and disinfection
Nasal cannula
Dispose of
Oxygen safety hoses
Dispose of
FireSafe check valve
Dispose of
Oxygen mask
Dispose of
Device inlet filter
Dispose of
9.3.3 Following a Technical Service
Component
Required action
aeroplus E
Cleaning and disinfection
Humidifier
Dispose of
Holder for humidifier
Cleaning and disinfection
Nasal cannula
Dispose of