Before starting cleaning the device must be
switched off and disconnected from the mains
supply.
1 Use a damp cloth to wipe down the exterior case
1 Unscrew the humidifier from the angle connection.
2 Unscrew the lid from the humidifier and pour out any
remaining water.
3 Clean the humidifier with clear, warm water.
4 Fill fresh water into the humidifier
5 Screw cover onto humidifier
6 Screw angle connection to humidifier
Sterile Water Prefilled Humidifiers
When using sterile water prefilled humidifier systems:
– Prefilled Humidifiers may not be cleaned and refilled.
– The old water flask must be disposed of.
– The information provided by the sterile water producer
(package insert) must be observed!
Cleaning nasal
cannulas, oxygen
hoses and masks
1 Disconnect the hose of the nasal cannula from aeroplus E.
2 Clean the nasal cannula in warm soapsuds. You may
alternatively use a weak acetic solution (10% vinegar, 90%
water).
3 Rinse the nasal cannula with lots of clear water.
4 Let the nasal cannula dry in air. The nasal cannula may
only be used again for the therapy after it has properly
dried.
Replacing the inlet
filter
1 Open the blue service lid in the back of the device.
2 Remove the old inlet filter by pulling it smoothly back.
3 Attach the new inlet filter.
4 Close the blue service lid.