aeroplus E
Maintenance
41
Maintenance activity
Description
Before starting cleaning the device must be
switched off and disconnected from the mains
supply.
1 Use a damp cloth to wipe down the exterior case
Cleaning the
humidifier
1 Unscrew the humidifier from the angle connection.
2 Unscrew the lid from the humidifier and pour out any
remaining water.
3 Clean the humidifier with clear, warm water.
4 Fill fresh water into the humidifier
5 Screw cover onto humidifier
6 Screw angle connection to humidifier
Sterile Water Prefilled Humidifiers
When using sterile water prefilled humidifier systems:
Prefilled Humidifiers may not be cleaned and refilled.
The old water flask must be disposed of.
The information provided by the sterile water producer
(package insert) must be observed!
Cleaning nasal
cannulas, oxygen
hoses and masks
1 Disconnect the hose of the nasal cannula from aeroplus E.
2 Clean the nasal cannula in warm soapsuds. You may
alternatively use a weak acetic solution (10% vinegar, 90%
water).
3 Rinse the nasal cannula with lots of clear water.
4 Let the nasal cannula dry in air. The nasal cannula may
only be used again for the therapy after it has properly
dried.
Replacing the inlet
filter
1 Open the blue service lid in the back of the device.
2 Remove the old inlet filter by pulling it smoothly back.
3 Attach the new inlet filter.
4 Close the blue service lid.