Kröber O2 - Kröber O2 Vers. 4.0
Taking into service
74 EN
7.3.1 Without external humidification
If the unit is to stand next to the user, it is started up as follows:
Fig. 8: Mains lead
1 Connect the Kröber O2 to mains.
ATTENTION!
The Kröber O2 is only designed
for operation on a 230 Volt,
50Hz mains network.
Fig. 9: Angular connection
2 The metal oxygen outlet is installed
using a wrench (metric size 17).
ATTENTION! Fire safety!
The metal oxygen outlet must be
installed at all times!
Fig. 10: Humidifier Max.-mark
3 Fill the humidifier up to the top
MAX.-mark (1).
ATTENTION! Sterility!
To ensure an adequate level of
sterility, use always boiled
water!
1