EN 75
Fig. 11: Humidifier assembly
4 Fasten the humidifier on the
angular connection (1).
5 Connect the 0.33m oxygen hose to
the humidifier outlet (2).
Fig. 12: Connection to nasal cannula
6 Plug the FireSafe Nozzle in oxygen
flow direction into the nasal
cannula.
ATTENTION!
Observe flow direction
indicated by the arrow!
ATTENTION!
For the best protection of the
patient, place the FireSafe
Nozzle in the oxygen tubing
routing as close to the patient
as possible!
7 Connect the other terminal to
the short oxygen tubing (0.33
m).
1
2
Nasal cannula
Oxygen hose