EN 91
9.3 Maintenance plan
Maintenance and cleaning work must be performed at regular intervals as specified in the
table below.
All below mentioned maintenance work may be carried out by the user. Eventually, a new
spare part has to be purchased beforehand. Please contact your service provider or the
manufacturer.
WARNING! Danger due to lack of authorization!
All maintenance work and in particular repairs not described in these operating
instructions may only be carried out by persons who have been authorised by
the manufacturer to do so in accordance with the instructions in the service
manual.
WARNING! Patient as operator!
Service and maintenance must not be carried out while the device is in use.
WARNING! Risk of infection!
When oxygen delivery accessories such as nasal cannulas are used together,
infection cannot be ruled out if the product is used by different users together.
Every user of the Kröber O2 should therefore use his own application set!
9.3.1 Clinical / Non-clinical Use
Component
Clinical Use
Non-clinical Use
KröberO2
Weekly cleaning and
disinfection
Weekly cleaning
Humidifier
Weekly cleaning and disinfection
replace annually
Mounting bracket for
humidifier
Weekly cleaning and
disinfection
Weekly cleaning
Nasal cannula
Weekly cleaning and
disinfection
replace monthly
Weekly cleaning
replace monthly
Oxygen safety tubing
Replace every 6 months
FireSafe check valve
Replace every 6 months