Kröber O2 - Kröber O2 Vers. 4.0
Maintenance
92 EN
Component
Clinical Use
Non-clinical Use
Oxygen mask
Weekly cleaning and
disinfection
replace monthly
Weekly cleaning
replace monthly
Coarse dust filter
replace monthly
Device inlet filter
replace annually / after 5000 operating hours
9.3.2 When changing patients
Component
KröberO2
Cleaning and disinfection
Humidifier
Dispose of
Mounting bracket for
humidifier
Cleaning and disinfection
Nasal cannula
Dispose of
Oxygen safety tubing
Dispose of
FireSafe check valve
Dispose of
Oxygen mask
Dispose of
Coarse dust filter
Dispose of
Device inlet filter
Dispose of
9.3.3 After technical service
Component
KröberO2
Cleaning and disinfection
Humidifier
Dispose of