INSTRUCTIONMANUAL
DIGITALCLAMPMETER
KEWSNAP
SERIES
KEW2040
 
600AACOnlyType
KEW2055
1000AAC/DCType
Features
Designedtomeetinternationalsafetystandards.
IEC61010-1.IEC61010-031:2002&IEC61010-2-032
MeasurementCategory(CAT.)IV600V
PollutionDegree2
Double molded main body provides comfortable
singlehandedgrip
DataHoldFunction
LCD Backlight function to facilitate working at
dimlylitsituations.(KEW2055only)
RELfunctiontoindicatemeasurementvariation.
(Current,voltage,Resistancemeasurement)
MIN/MAX function enables easy reading of min &
maxvalueduringmeasurement.
WithContinuity&DiodeCheckFunction
NCV
Non Contact Voltage
Function for wiring
check.
600Vinputprotection.
Sleepfunctiontoextendbatterylife.
WithBarGraph,6039counts
2.SafetyWarnings
This instrument has been designed, manufactured and
testedaccordingtoIEC61010:Safetyrequirementsfor
Electronic Measuring apparatus, and delivered in
the best condition after passed the inspection. This
instruction manual contains warnings and safety
rules which must be observed by the user to ensure
safeoperationoftheinstrumentand retain it in safe
condition.
Therefore, read through these operating instructions
beforeusingtheinstrument.
WARNING
Read through and understand the instructions
contained in this manual before using the
instrument.
Keep the manual at hand to enable quick
referencewhenevernecessary.
Theinstrumentistobeusedonlyinitsintended
applications.
Understa n d a n d f o l l o w a l l t h e s a f e t y
instructionscontainedinthemanual.
It is essential that the above instructions are
adheredto.
Failure to follow the above instructions may
causeinjury,instrumentdamageand/ordamage
toequipmentundertest.
Thesymbol
indicatedontheinstrumentmeans that
theusermustrefertotherelatedpartsinthemanual
forsafeoperationof the instrument.Itisessential to
read the instructions wherever the
symbol appears
inthemanual.

DANGERisreservedforconditions andactions
thatarelikelytocauseseriousorfatalinjury.

WARNINGisreservedforconditionsandactions
thatcancauseseriousorfatalinjury.

CAUTIONisreservedforconditionsandactions
thatcancauseinjuryorinstrumentdamage.
Marks listed in the table below are used on this
instrument.
Usermustrefertothemanual.
Instrumentwithdoubleorreinforcedinsulation
Indicates that this instrument can clamp on
bare conductors when measuring a voltage
corresponding to the applicable measurement
category,whichismarkednexttothissymbol.
AC
DC
AC&DC
DANGER
Nevermakemeasurementonacircuitinwhich
voltageoverAC600Vexists.
Do not attempt to make measurement in the
presence of flammable gasses. Otherwise, the
use of the instrument may cause sparking,
whichcanleadtoanexplosion.
Transformerjawtipsare designednot toshort
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precaution should be taken to minimize the
possibilityofshorting.
Never attempt to use the instrument if its
surfaceoryourhandiswet.
Donotexceedthemaximumallowableinputof
anymeasuringrange.
Never o pen t he Bat ter y cov er d uri ng a
measurement.
Theinstrumentistobeusedonlyinitsintended
applications or conditions. Otherwise, safety
functions equipped with the instrument doesn
t
work,andinstrumentdamageorseriouspersonal
injurymaybecaused.
WARNING
Never attempt to make measurement if any
abnormal conditions, such as broken case
and exposed metal parts are found on the
instrument.
Do not rotate the Function Switch while the
testleadsarebeingconnected.
Do not install substitute parts or make any
modification to the instrument. For repair or
re-calibration, return the instrument to your
localdistributorfromwhereitwaspurchased.
Do not try to replace the batteries if the
surfaceoftheinstrumentiswet.
Disconnect all the cords and cables from the
objectundertestandpowerofftheinstrument
before opening the Battery Cover for Battery
replacement.
Verify proper operation on a known source
before use or taking action as a result of the
indicationoftheinstrument.
Use appropriate personal protective equipment
suchasinsulatinggloves,Insulatingboots,and
safetyglasses.
CAUTION
Set the Function Switch to an appropriate
positionbeforestartingmeasurement.
Firmlyinsertthetestleads.
Disconnect the test leads from the instrument
forcurrentmeasurement.
Donotexposethe instrumenttothedirectsun,
hightemperatureandhumidityordewfall.
Altitude 2000m or less. Appropriate operating
temperatureiswithin0
and40
.
Thisinstrumentisn
tdust&waterproofed.Keep
awayfromdustandwater.
Be sure to power off the instrument after use.
When the instrument will not be in use for a
longperiod,placeitinstorageafter removing
thebatteries.
U se a c lot h dip ped in w ate r o r neu tra l
detergent for cleaning the instrument. Do not
useabrasivesorsolvents.
Measurementcategories
(Over-voltagecategories)
To ensure safe operation of measuring instruments,
IEC61010 establishes safety standards for various
electrical environments, categorized as CAT.I to
CAT.IV, and called measurement categories. Higher-
nu mbered c at egories co rrespond to e le ctrical
environment s with grea ter momentary energy ,
so a measuring instrument designed for CAT.III
environments can endure greater momentary energy
thanonedesignedforCAT.II.
CAT.Ⅰ
:Secondary electricalcircuitsconnectedtoan
ACelectricaloutletthroughatransformeror
similardevice.
CAT.Ⅱ
:Primary electrical circuits of equipment
connected to an AC electrical outlet by a
powercord.
CAT.Ⅲ
:Primary electrical circuits of the equipment
connected directly to the distribution panel,
and feeders from the distribution panel to
outlets.
CAT.Ⅳ
:The circuit from the service drop to the
service entrance, and to the power meter
and primary over current protection device
(distributionpanel).
3.Specification
3-1.Measuringrange&accuracy
(accuracyguaranteedat23
℃±
5
.humidity45
85%)
ACCurrent600A,1000AFunction
Function
Measuring
Range
Accuracy
KEW2040 KEW2055
600A 0-600.0A
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
1000A 0-1000A
N/A
DCCurrent600A,1000AFunction
Function
Measuring
Range
Accuracy
KEW2040 KEW2055
600A
0-600.0A
N/A
±
1.5%rdg
±
5dgt
1000A
0-1000A
N/A
ACVoltageFunction

Auto-ranging,Inputimpedance:approx.10M
Ω
Range
Measuring
Range
Accuracy
KEW2040 KEW2055
6/60/600V 1-600.0V
±
1.3%rdg
±
4dgt(50/60Hz)
±
3.0%rdg
±
5dgt(40
400Hz)
DCVoltageFunction

Auto-ranging,Inputimpedance:approx.10M
Ω
Range
Measuring
Range
Accuracy
KEW2040 KEW2055
600mV/6/
60/600V
0-600.0V
±
1.0%rdg
±
3dgt
Resistance
Continuity/DiodeCheck
Function
Range
Measuring
Range
Accuracy
KEW2040 KEW2055
600
Ω
/6k/
60k/600k
Ω
6M
Ω
0-6M
Ω
±
1.0%rdg
±
5dgt
60M
Ω
6.00M-
60.00M
Ω
±
5%rdg
±
8dgt
ContBuzzer 0-600.0
Ω
Thresholdvalue:60
Ω±
30
Ω
Diode
Testvoltage:0-2V
Frequ enc y/ D UTY F uncti on
Auto- ran gin g f or
Frequency
Range
Measuring
Range
Accuracy
KEW2040 KEW2055
ACA 40Hz
400Hz
±
0.5%rdg
±
5dgt
ACV 1Hz
10kHz
0.1-99.9%
(Pulsewidth/Pulseperiod)
±
2.5%rdg
±
5dgt
Note:Measurableinputsare:40Vrms@ACVor
60Arms@AC600A,350A@AC1000ARange
3-2.GeneralSpecification
Modeofoperation
:
Δ
∑mode
Display
:max.6039counts

     
(Frequency:9999)&Bargraph
Over-range indication : "OL" displayed when exceeding
the measuring range. (except for AC/DCV and 1000A
Function)
Rangeswitching :
Auto-ranging/Voltage,ResistanceRange
Singlerange/Continuity,Diodecheckand
DUTY
Samplerate
  
:threetimespersecond
Functionalconstruction:
OFF/ACA/ACV/DCV/
Ω
KEW2040
OFF/ACA/DCA/ACV/DCV/
Ω
KEW2055
Keys :
SELECT(AC/DCswitching&/
Ω
/ / ),REL
Δ
,
Hz/DUTY,MIN/MAX,HOLD/BackLight(KEW2055)
Powersource :DC3V/R03(UM-4)x2pcs
Lowbatterywarning
:
" "markisdisplayedat
   
2.4V
±
0.15Vorless.
Temperature & humidity : 23
℃±
5
, relative humidity
accuracyguaranteed85%orless(nocondensation)
Operatingtemperature:0
40
,relativehumidity
&humidityrange85%orless(nocondensation)
Storagetemperature :-20
60
,relativehumidity
&humidityrange85%orless(nocondensation)
Currentconsumption:approx.12mA
SleepFunction:Automaticallypoweredoffinabout
15minafterthelastFunctionswitchoperation.
Press any key or rotate the Function Switch from
OFFtoanypositiontoexitfromtheSleepstate.
ApplicableStandards
IEC61010-1:2001
MeasurementCAT.IV600VPollutiondegree2
IEC61010-031:2002,IEC61010-2-032
EMC
EN61326
EN55022
EN61000-4-2(performancecriterionB)
EN61000-4-3(performancecriterionB)
OverloadProtection
CurrentRange :720AAC/10sec@KEW2040
1200AAC/DC/10sec@KEW2055
VoltageRange :720VAC/DC/10sec
ResistanceRange :600VAC/DC/10sec
WithstandVoltage
6880VAC
TRMS50/60Hz
/5sec
(betweenJawsandelectricalcircuit/betweeninternal
circuitandenclosure)
InsulationResistance:10M
Ω
ormore/1000V
betweenelectricalcircuitandenclosure
Conductorsize
KEW2040:approx.33mm
KEW2055:approx.40mm
Dimension
approx.254(L)
×
82(W)
×
36(D)mm/KEW2055
approx.243(L)
×
77(W)
×
36(D)mm/KEW2040
Weight:approx.:300g@KEW2040
:310g@KEW2055
Accessories
TestLeads Model7066/1set
BatteryR03
UM-4
/2pcs
InstructionmanualEnglish,Japanese/1pce
CarriyngCaseModel9094/1pcs
3-3.FunctionKeys
The"
"markshowsavailablefunctionateachRange.
DCCurrentMeasurement(KEW2055only)
HOLD SELECT ZERO
Hz/
DUTY
MAX/
MIN
ACA
ACV
-
DCA
-
DCV
-
-
Ω
-
-
- - -
-
- - -
4.Preparationformeasurement
4-1.CheckingBatteryVoltage
Set the Function Switch to any position other than
"
OFF
".Whenthedisplay isclearwithout"
BATT
"mark,
showing, battery voltage is enough. When the display
is blank or
"
BATT
" mark is indicated, replace the
batteriesaccordingtoSection7,BatteryReplacement.
CAUTION
The Sleep feature automatically powers the
instrument off in about 15 min after the last
switch or key operation. Therefore, the display
may beblankevenwiththeFunctionSwitchset
to a position other than "
OFF
". To operate the
instrument in this case, turn the switch back to
the "
OFF
" position, then to any other position,
orpressanykey.Replacethebatteriesifnothing
wasdisplayedafteraboveoperations.
4-2.CheckingSwitchSetting&Operation
Confirm the Function Switch is set to the correct
posit ion , t he i nst rumen t i s set to the c orrec t
measurementmode,
and the Data hold function is disabled. Otherwise,
desiredmeasurementcannotbemade.
5.Measurement
5-1.ACCurrentMeasurement
DANGER
Never make measurementon a circuitin which
voltage over AC600V exists to avoid getting
electricalshock.
Transformer jaw tips are designed not to short
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precaution should be taken to minimize the
possibilityofshorting.
Do not make measurement with the Battery
Coverremoved.
Disconnectthetestleadsfromtheinstrumentfor
currentmeasurement.
(1)Set the Function Switch to "
600A
" or "
1000A
"
position.(onKEW2040,only"600A"isavailable)
AChasbeenselectedbydefault;presstheSELECT
key, when DC has been selected, to change it to
AC.ACmark is displayed attheupper left onthe
display.(KEW2055only)
(2)Press the trigger to open the transformer jaws
andclampthemontotheoneconductorundertest,
then take the reading on the display. Pressing the
"
Hz/DUTY
"Keyswitchestheindicationinfollowing
sequence.
ACCurrent
Hz
DUTY
Hz/DUTY
Function requires 60A or more at AC600A
Rangeand350AormoreatAC1000Arange.
CAUTION
MaxconductorsizeforKEW2040isapproxdia.
33mm and for KEW2055 is approx dia. 40mm.
During current measurement, keep the trans
former jaws fully closed. Other wise, accurate
measurementscannotbetaken.
5-2.DCCurrentMeasurement(KEW2055only)
DANGER
Nevermakemeasurementonacircuitinwhich
voltage over DC600V exists to avoid getting
electricalshock.
Do not make measurement with the Battery
Coverremoved.
(1)Set the Function Switch to "
600A
" or "
1000A
"
position. AC has been selected by default; press
the SELECT key, when AC has been selected, to
changeittoDC.
DC mark is displayed at the upper left on the
Transformer
Question and Answer IconNeed help?

Do you have a question about the KYORITSU KEW SNAP Series KEW2040 and is the answer not in the manual?

KYORITSU KEW SNAP Series KEW2040 Specifications

General IconGeneral
ModelKEW2040
TypeClamp Meter
AC Voltage Measurement600 V
DC Voltage Measurement600 V
DisplayDigital
Battery2 x LR03 (AAA)
Safety StandardIEC 61010-1 CAT III 300V / CAT II 600V
WeightApprox. 220g (including batteries)

Summary

Features

Safety Warnings

Specifications

Measuring Range & Accuracy

Details AC/DC current, voltage, resistance, and frequency ranges and their accuracy specifications.

General Specification

Covers display, range switching, sample rate, functional construction, power source, and environmental conditions.

Function Keys

Outlines the available functions for each key (HOLD, SELECT, ZERO, Hz/DUTY, MIN/MAX) across different ranges.

Measurement

AC Current Measurement

Step-by-step guide for measuring AC current using the clamp meter, including safety warnings.

DC Current Measurement

Detailed procedure for measuring DC current, emphasizing safety precautions and key operations.

AC Voltage Measurement

Instructions for measuring AC voltage, including the use of Hz/DUTY function and safety notes.

DC Voltage Measurement

Procedure for measuring DC voltage, highlighting connection methods and safety precautions.

Resistance/ Cont/ Diode Measurement

Guide for measuring resistance, continuity, and diodes, with specific instructions for each mode.

Other Functions

Sleep Function

Explains the automatic power-off feature to conserve battery and how to exit sleep mode.

HOLD Key

Describes the Data Hold function to freeze readings and the Backlight ON/OFF feature.

NCV Function

Details the Non-Contact Voltage detection feature and its operational indicators.

MIN/MAX Function

Instructions for capturing minimum and maximum values during measurements.

ZERO Function

Explains the zero adjustment function for current measurements and relative value indication.

Battery Replacement

Maintenance

Related product manuals