EasyManuals Logo

Labconco FlaskScrubber Vantage Series User Manual

Default Icon
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
Chapter 7: Troubleshooting
Product Service 1-800-522-7658 55
PROBLEM CAUSE CORRECTIVE ACTION
Defective controller Check output from PCB to both valves and DI
pump.
Replace PCB. Contact Labconco Product
Service.
Holes in filter screen
too small
Contact Labconco Product Service.
Defective grommet
where hose connects
to fill level switch
Replace grommet.
Low Temp Alarm
Heater inoperative Check electrical connections.
Check voltage to heater.
Check resistance of heater. If open circuit,
replace heater.
Defective controller Check output from PCB to heater.
Replace PCB. Contact Labconco Product Service.
Defective
temperature probe
Check electrical connections.
Contact Labconco Product Service. Replace if
necessary.
Defective high
temperature cut out
switch
Check resistance of safety cut out switch. If open
circuit at room temperature, replace switch.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco FlaskScrubber Vantage Series and is the answer not in the manual?

Labconco FlaskScrubber Vantage Series Specifications

General IconGeneral
BrandLabconco
ModelFlaskScrubber Vantage Series
CategoryWasher
LanguageEnglish

Related product manuals