This guide provides site preparation information for the Ech 650 Series Liquid Handler. This includes models 650, 655, 650T
and 655T. The topics below describe the site preparation requirements and instrument specifications for the Echo Liquid
Handlers.
Pre-Installation Requirements
l Site Requirements
l User Supplied Equipment Requirements
l Communication Requirements
l Accommodating Your Echo Liquid Handler
l Noise Ratings
l Power Ratings
What You Will Be Receiving
l Shipping Dimensions
l Echo Liquid Handler Components
l Chiller
l Junair Compressor (Optional and sold separately by Labcyte)
l Vacuum Pump (Optional and sold separately by Labcyte)
l Client PC (Sold separately by Labcyte)
Pre-Installation Requirements Verification Form
LABCYTE INC.
SITE PREP GUIDE | Echo® 650 Series Liquid Handler
© 2019 LABCYTE INC. All rights reserved.
QSG-2019-1
FEB 2019
Echo® 650 Series Liquid Handler
(Models: 650, 655, 650T, 655T)
SITE PREP GUIDE
Server Version 3.1 | FEBRUARY 2019
Question and Answer IconNeed help?

Do you have a question about the Labcyte Echo 650 Series and is the answer not in the manual?

Labcyte Echo 650 Series Specifications

General IconGeneral
TypeAcoustic liquid handler
TechnologyAcoustic Droplet Ejection (ADE)
Plate Formats96-, 384-, and 1536-well plates
Power Requirements100-240 VAC, 50/60 Hz
PrecisionCV < 5% at 10 µL
Supported Microplate FormatsSBS standard microplates
Source Plate CapacityUp to 6 plates
ApplicationsGenomics, drug discovery, proteomics

Summary

Pre-Installation Requirements

Site Requirements

General requirements for the installation site.

User Supplied Equipment Requirements

Equipment the user must provide for the installation.

Communication Requirements

Network and communication protocols needed for the system.

Accommodating Your Echo Liquid Handler

Space and layout considerations for the instrument.

Noise and Power Ratings

Sound level and electrical power consumption specifications.

What You Will Be Receiving

Shipping and Components

Physical dimensions, weight, and list of included parts.

Optional Equipment

Details on optional chiller, compressor, and vacuum pump.

Client PC

Specifications for the required client computer.

Pre-Installation Requirements Verification Form

Site Requirements

Environment and Power

Specifies environment, AC power options, and socket needs.

Compressed Air Source

Defines requirements for clean, filtered, oil-free, dry air pressure and flow.

User Supplied Equipment Requirements

Hardware Fittings

Specifies fittings for compressed air and vacuum lines.

Client PC Specifications

Lists minimum OS, CPU, RAM, storage, and display requirements.

Communication Requirements

Server, Network, and User Account

Specifies server version, network protocols, and user account access.

Accommodating Your Echo Liquid Handler

Space and Clearance

Details instrument dimensions, working envelope, and space for accessories.

Noise Ratings

Sound Level

Specifies the noise level of the Echo Liquid Handler in dBA.

Power Ratings

Power Consumption

Details the power (W) and heat rating of the Echo Liquid Handler.

AC Power, Configuration, and Fuses

Electrical Configurations

Table detailing voltage, frequency, power, and fuse requirements.

Chiller Specifications

Power, Control, and Conditions

Details power supply, temp control, and operating conditions.

Junair Compressor Specifications

Compressor Performance

Table of specifications for the Junair OF302-4B compressor.

Vacuum Pump Specifications

Source Requirements & Vacuubrand Specs

Defines source needs and details Vacuubrand MZ 2S-NT specs.

Client PC Specifications

System Requirements

Covers OS, CPU, RAM, storage, video, and network.

Accessories and Software

Lists included peripherals and optional software applications.

Pre-Installation Requirements Verification Form

Verification Checklist

Checklist to confirm all pre-installation requirements are met prior to scheduling installation.