Function Range Resolution
Accuracy
% of measured value (rdg)
+ least significant digits
DC voltage
600.0 mV 0.1 mV ± (0,5% rdg ± 8 digits)
6.000 V 1 mV
± (0,8% rdg ± 5 digits)60.00 V 10 mV
600.0 V 0.1 V
1000 V 1 V ± (1,0% rdg ± 3 digits)
AC+DC voltage
Bandwidth: 50-400 Hz
6.000 V 1 mV
± (1,5% rdg ± 20 digits)60.00 V 10 mV
600.0 V 0.1 V
1000 V 1 V ± (1,5% rdg ± 5 digits)
AC/DC voltage (LOW Z)
Bandwidth: 50-400 Hz
6.000V 1 mV
± (3,0% rdg ± 30 digits)60.00 V 10 mV
600.0 V 0.1 V
1000 V 1 V ± (3,0% rdg ± 5 digits)
Resistance
600.0 Ω 0.1 Ω
± (1,5% rdg ± 5 digits)
6.000 kΩ 1 Ω
60.00 kΩ 10 Ω
600.0 kΩ 100 Ω
6.000 MΩ 1 kΩ
60.00 MΩ 10 kΩ ± (2,0% rdg ± 10 digits)
Capacitance
60.00 nF 10 pF ± (5,0% rdg ± 35 digits)
600.0 nF 100 pF
± (3,0% rdg ± 5 digits)
6.000 µF 0.001 µF
60.00 µF 0.01 µF
600.0 µF 0.1 µF
6000 µF 1 µF ± (5,0% rdg ± 5 digits)
Frequency
9.999 Hz 0.001 Hz
± (1,0% rdg ± 5 digits)
99.99 Hz 0.01 Hz
999.9 Hz 0.1 Hz
9.999 kHz 1 Hz
Pulse width: 100 µs ... 100 ms
Frequency: 10 Hz ... 1 kHz
Sensitivity: > 8 V RMS
Duty cycle
20% … 80% 0.1% ± (1,2% rdg ± 2 digits)
Pulse width: 100 µs ... 100 ms
Frequency: 10 Hz ... 1 kHz
Sensitivity: > 8 V RMS
Temperature
-20 … 760°C 0.1~1°C ± (1% rdg ± 5°C)
-4 … 1400°F 0.1~1°F ± (1% rdg ± 9°F)
EN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LaserLiner MultiMeter XP and is the answer not in the manual?

LaserLiner MultiMeter XP Specifications

General IconGeneral
BrandLaserLiner
ModelMultiMeter XP
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals