Temperature Range | -20°C to 200°C |
---|---|
Temperature Stability | ±0.01°C |
Refrigerant | R134a |
Power Supply | 230 V, 50/60 Hz |
Bath Volume | 2 l |
Heating Power | 2 kW |
Cooling Capacity | 0.2 kW at 20°C |
Introduces the LAUDA Puridest Distillation Apparatus models.
Produces highly-pure, bacteria and pyrogen-free distillate meeting pharmacopeia standards.
Identifies key components like main switch, pilot lamps, and withdrawal tap.
Advices checking package contents and contacting the supplier for damages or complaints.
Describes the apparatus's function for boiling water and producing distillate.
Lists restrictions on use, including not for medical devices or explosive environments.
Details the warranty claim process and required information for defective units.
Explains warning symbols for hot liquids, electrical voltage, and general warnings.
Provides guidance on safe transport, lifting, and placement on stable, watertight surfaces.
Details how to use keyholes for wall mounting and the required fixing material.
States the unit is not suitable for potentially explosive surroundings.
Specifies requirements for connection to the mains supply, including shock-proof sockets and earth conductor.
Locates water connections on the right-hand side of the unit.
Explains how to withdraw distillate using the front tap and connecting a hose.
Describes connecting a hose to drain the boiler for cleaning or maintenance.
Details connecting a hose to the cooling water outlet for drainage.
Explains connecting the tap water inlet via a solenoid valve and pressure hose.
Instructions for filling the apparatus with water and re-assembling lids.
How to start the apparatus, including discarding initial distillate.
Procedure for switching off the unit for longer periods and emptying.
Explains the automatic sequence of operation, including pilot lamps and water level control.
Emphasizes cooling down the unit and disconnecting power before maintenance.
Provides detailed steps for descaling using a descaling agent and rinsing.
Addresses issues like foam formation causing the "Clean" pilot lamp to light up.
Explains foam formation causes and the resulting unit shutdown.
Discusses causes for back water in the boiler and the "Clean" lamp indication.
Steps to reset the low water cut-off and re-initiate the unit.
Advice on handling, cleaning surfaces, and servicing by qualified personnel.
Explains LAUDA-GFL's responsibility for environmentally sound handling and recycling of units.
Provides exterior dimensions, storage tank size, and distillation capacity for models.
Details distillate quality parameters like conductivity and purity.
Lists cooling water requirements, pressure, electrical connection, and ambient conditions.
Provides exterior dimensions, storage tank size, and distillation capacity for models.
Details electrical connections, mains supply, fuses, and operating conditions.
Lists and defines components used in the circuit diagrams.
Presents the electrical circuit diagram for the PD 2 R model.
Presents the electrical circuit diagram for the PD 4 R model.
Shows the electrical circuit for the PD 8 R model with 230V/1~ connection.
Presents the electrical circuit for PD 8 R / PD 12 R models with 220V/3~ connection.
Shows the electrical circuit for PD 8 R / PD 12 R models with 400V/3~ connection.
Explains the colour coding of wires for mains connection cables.
Details compatibility for connecting PD 2/4 R and PD 8/12 R to different mains supplies.
Lists power consumption and required mains fuses for each model.
Lists components like sockets, plugs, and main switches for mains connection.
Example connection for PD 2 R / PD 4 R to 220 V / 3 ~ mains supply.
Example connection for PD 2 R / PD 4 R to 400 V / 3 ~ mains supply.
Example connection for PD 8 R (230V/1~) to 220 V / 3 ~ mains.
Example connection for PD 8 R (230V/1~) to 400 V / 3 ~ mains.
Example connection for PD 8 R / PD 12 R to 220 V / 3 ~ mains via switch.
Example connection for PD 8 R / PD 12 R to 220 V / 3 ~ mains via CEE plug.
Example connection for PD 8 R / PD 12 R to 400 V / 3 ~ mains via switch.
Example connection for PD 8 R / PD 12 R to 400 V / 3 ~ mains via CEE plug.
Lists spare parts with part numbers and descriptions for various components.
Describes options for separate water supply, with and without solenoid valve.
Details Dechlorite Filter and Phosphate cartridge for water pre-treatment.
Describes pre-filters for tap water cleaning and wall brackets for filter mounting.
Details hose sets for water connections and level control switches for external tanks.
Declares conformity of Puridest models with EC Directives and lists technical data.
Lists the standards applied for Low Voltage, EMC, and RoHS Directives.