Questions and Answers

  • B
    Brian StricklandAug 14, 2025
    What to do if my LEYBOLD D10E Water Pump does not start?
    • L
      Leon PughAug 15, 2025
      If your LEYBOLD Water Pump isn't starting, several issues could be the cause. First, check the wiring and repair it if faulty. For three-phase motors, ensure the motor protection switch is properly set up. Also, verify that the power supply voltage matches the motor's rating. Other potential causes include a faulty motor (exchange it), oil temperature being less than 12 °C (warm up the pump and oil or use a different oil), oil being too thick (change the oil), a clogged exhaust filter or line (exchange the filter or clean the line), or a seized pump (repair the pump).
  • W
    william40Aug 18, 2025
    Why is my LEYBOLD D10E Water Pump getting hotter than usual?
    • M
      Mandy KimAug 18, 2025
      If your LEYBOLD Water Pump is overheating, several factors could be responsible. Check if the supply of cooling air is obstructed and ensure the pump is properly installed. High ambient temperature or process gas that is too hot can also cause overheating, so ensure the pump is properly installed and change the process if needed. Insufficient oil levels or unsuitable oil types could be the cause, so fill in or change the oil. Other potential issues include obstructed oil circulation (clean or repair the oil channels), a clogged exhaust filter or line (change the filter or clean the line), a faulty exhaust valve (repair the valve), or deviating mains supply voltage (check the power supply rating of the motor and the mains supply).
  • J
    Joann HillAug 21, 2025
    What to do if my LEYBOLD Water Pump is extremely loud?
    • H
      hcarsonAug 21, 2025
      If your LEYBOLD Water Pump is making excessive noise, start by checking the oil level and top it up if it's too low. High intake pressure can also cause noise, so reduce it if necessary. Other potential causes include a clogged internal demister (clean or replace it), worn coupling bushings (replace them), damaged vanes or bearings (repair the pump), or a faulty motor bearing (exchange the motor bearing).
  • E
    Ernest MorrisAug 25, 2025
    Why LEYBOLD D10E does not attain its ultimate pressure?
    • T
      Teresa WilsonAug 25, 2025
      If your LEYBOLD Water Pump isn't reaching its ultimate pressure, consider these possible causes. Ensure you're using the right method of measurement or measuring instrument. Check for external leaks by measuring the pressure directly at the intake port of the pump. Other potential issues include a dirty internal demister or faulty exhaust valve (repair the pump or valve), unsuitable oil (change and possibly degas it), a dirty or leaky intake line (clean or exchange it), or the pump being too small (check process data and possibly change the pump).
  • T
    Timothy JohnsonAug 28, 2025
    What to do if pumping speed of LEYBOLD D10E Water Pump is too low?
    • U
      ustoneAug 28, 2025
      If the pumping speed of your LEYBOLD Water Pump is too low, check if the dirt trap in the intake line is clogged and clean it if necessary. Additionally, a clogged exhaust filter can reduce pumping speed, so exchange the filter element. Finally, ensure that the connection lines are sufficiently wide and short.
  • T
    Thomas CampbellAug 31, 2025
    What to do if there is oil in the intake line of my LEYBOLD D10E?
    • R
      Robert PerryAug 31, 2025
      If you find oil in the intake line or vacuum chamber of your LEYBOLD Water Pump, it could be coming from the system itself, so check the system. Another possibility is a blocked tandem valve, in that case clean or repair it.
  • S
    Samuel HowardSep 3, 2025
    What to do if after switching off LEYBOLD D10E Water Pump the pressure increases too fast?
    • D
      Dominic SharpSep 3, 2025
      If the pressure in the system increases too fast after switching off your LEYBOLD Water Pump, it could indicate a system leak, so check the system. Other potential causes include a faulty tandem valve, inner shaft seal, or exhaust/bypass valve. In such cases, repair the valve(s) or exchange the shaft seal.
  • Z
    zthomasSep 7, 2025
    Why is the oil in my LEYBOLD D10E Water Pump turbid?
    • W
      wcobbSep 7, 2025
      If the oil in your LEYBOLD Water Pump is turbid, it's likely due to condensation. You should degas the oil or change the oil and clean the pump.
  • B
    Bryan BrowningSep 10, 2025
    Why is there an oil leak at the side of the motor of my LEYBOLD D10E?
    • C
      Charles LeeSep 10, 2025
      If you notice an oil leak at the side of the motor of your LEYBOLD Water Pump, it's likely due to a damaged or worn-out outer shaft seal. You should exchange the shaft seal.

Summary

1 Description

1.1 Principle of Operation

Explains how the dual-stage rotary vane vacuum pump functions, including gas compression and oil lubrication.

1.4 Technical Data

Details the performance specifications, dimensions, and electrical data for the TRIVAC E models.

2 Operation

2.1 Installation

Guides on the proper placement and mounting of the TRIVAC E vacuum pump for optimal operation.

2.2 Connection to the System

Instructions for connecting the vacuum pump to the process system, including intake and exhaust lines.

2.3 Electrical Connection

Details on how to safely and correctly connect the TRIVAC E pumps to the electrical power supply.

2.5 Pump Operation Modes

Guidance on operating the pump for various gas types, including temperature and ballast considerations.

2.5.2 Pumping Condensable Gases

Guidance on pumping condensable gases and vapours, emphasizing gas ballast and operating temperature.

3 Checking; Maintenance

3.1 Oil Level and Quality Check

Procedures for checking and assessing the condition of the pump's oil.

3.2 Oil Change and Disposal

Step-by-step guide for changing pump oil and compliant disposal of used materials.

3.5 Electric Motor Maintenance

Instructions for disassembling and assembling the electric motor for maintenance.

4 Maintenance Plan (Recommendation)

5 Maintenance and Repair Kits

6 Troubleshooting

Pump Does Not Start Up

Identifies causes and solutions for the pump failing to start, such as wiring or motor issues.

Pump Does Not Attain Ultimate Pressure

Troubleshooting for issues related to the pump not reaching its specified ultimate pressure.

Low Pumping Speed

Addresses causes and remedies for reduced pumping speed, like clogged traps or filters.

Pressure Increase After Pump Shutdown

Explains why pressure rises after shutdown and how to fix it, e.g., faulty valves.

LEYBOLD D10E Specifications

General IconGeneral
BrandLEYBOLD
ModelD10E
CategoryWater Pump
LanguageEnglish