DETAILED INFORMATION ABOUT YOUR SYSTEM
System Accuracy
Diabetes experts have suggested that glucose meters should
agree within 15mg/dL of a laboratory method when the
glucose concentration is lower than 100mg/dL, and within
15% of a laboratory method when the glucose concentration
is 100mg/dL or higher. Samples from 100 patients were
tested using both the OneTouchSelect Plus Simple
®
System
and the YSI2300 Glucose Analyzer laboratory instrument.
System Accuracy Results for Glucose
Concentrations <100mg/dL
Percent (and number) of meter results that match the
laboratory test
Within ±5mg/dL Within ±10mg/dL Within ±15mg/dL
61.3% (103/168) 94.6% (159/168) 99.4% (167/168)
System Accuracy Results for Glucose
Concentrations Between 39.9mg/dL
and 451mg/dL
Percent (and number) of meter results that match the
laboratory test
Within ±15mg/dL or ±15%
99.2% (595/600)
NOTE: Where 39.9mg/dL represents the lowest glucose
reference value and 451mg/dL represents the highest
glucose reference value (YSI value).
System Accuracy Results for Glucose
Concentrations ≥100mg/dL
Percent (and number) of meter results that match the
laboratory test
Within ±5% Within ±10% Within ±15%
69.0% (298/432) 95.4% (412/432) 99.1% (428/432)