42
DETAILED INFORMATION ABOUT YOUR SYSTEM
Precision
Within Run Precision (300 Venous Blood Samples Tested per
Glucose Level)
Data generated using the OneTouchSelect Plus Simple
®
Meter.
Target
Glucose
(mg/dL)
Mean Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coecient of
Variation (%)
25 25.43 1.50 5.88
40 40.33 1.56 3.86
65 63.01 2.11 3.35
120 117.43 3.07 2.61
200 196.55 4.42 2.25
350 349.25 7.83 2.24
560 562.88 11.61 2.06
Results show that the greatest variability observed between test
strips when tested with blood is 2.11mg/dL SD or less at glucose
levels less than 100mg/dL, or 2.61% CV or less at glucose levels
at 100mg/dL or above.
Total Precision (600 Control Solution Tests per
Glucose Level)
Data generated using the OneTouchSelect Plus Simple
®
Meter.
Glucose Level
Ranges
(mg/dL)
Mean Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coecient of
Variation (%)
20 - 30 25.74 1.13 4.41
30 - 50 46.19 1.29 2.80
96 - 144 111.93 2.27 2.03
280 - 420 362.94 6.09 1.68
420 - 600 546.34 10.27 1.88