EasyManua.ls Logo

Lifescan OneTouch Select Plus Simple - Page 48

Lifescan OneTouch Select Plus Simple
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
DETAILED INFORMATION ABOUT YOUR SYSTEM
Precision
Within Run Precision (300 Venous Blood Samples Tested per
Glucose Level)
Data generated using the OneTouchSelect Plus Simple
®
Meter.
Target
Glucose
(mg/dL)
Mean Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coecient of
Variation (%)
25 25.43 1.50 5.88
40 40.33 1.56 3.86
65 63.01 2.11 3.35
120 117.43 3.07 2.61
200 196.55 4.42 2.25
350 349.25 7.83 2.24
560 562.88 11.61 2.06
Results show that the greatest variability observed between test
strips when tested with blood is 2.11mg/dL SD or less at glucose
levels less than 100mg/dL, or 2.61% CV or less at glucose levels
at 100mg/dL or above.
Total Precision (600 Control Solution Tests per
Glucose Level)
Data generated using the OneTouchSelect Plus Simple
®
Meter.
Glucose Level
Ranges
(mg/dL)
Mean Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coecient of
Variation (%)
20 - 30 25.74 1.13 4.41
30 - 50 46.19 1.29 2.80
96 - 144 111.93 2.27 2.03
280 - 420 362.94 6.09 1.68
420 - 600 546.34 10.27 1.88