b-4 Recommended communication settings on LTS-200.
Recommended settings for “COMM SETTING SELECTION” on LTS-200 special
function is as follows.
LTS-200 → PC(DL-01) to “SEND RECT. DATA”
1. BAUD RATE: 1200
2. DATA LENGTH: 8
3. PARITY BITS: NIL
4. STOP BITS: 1
5. SIGNAL CONTROL: OFF
6. XON/XOFF: OFF
7. PROTOCOL: OFF for “CSV” format, ON for “DC1 (DC-1Z)” format
8. RECORD DELIMETER: CR
Subsequent items: As you like.
LTS-200 → PC(DL-01) to “SEND POLAR DATA”
1. BAUD RATE: 1200-9600
2. DATA LENGTH: 8
3. PARITY BITS: NIL
4. STOP BITS: 1
5. SIGNAL CONTROL: OFF
6. XON/XOFF: OFF
7. ROTATION: As you like.
PC(DL-01) → LTS-200 to “RECEIVE RECT. DATA”
1. BAUD RATE: 1200
2. DATA LENGTH: 8
3. PARITY BITS: NIL
4. STOP BITS: 1
5. SIGNAL CONTROL: OFF
6. XON/XOFF: OFF
7. PROTOCOL: OFF for “CSV” format, ON for “DC1 (DC-1Z)” format
8. RECORD DELIMETER: CR
9. Subsequent items: As you like.
Please note that these settings should be common with DL-01's.
b-5 Recommended communication settings on DL-01.
To configure DL-01 Communication setting, please read “Configuring the software”
in the Help topics of DL-01 and select values as follows:
37

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Linertec LTS-205N and is the answer not in the manual?

Linertec LTS-205N Specifications

General IconGeneral
BrandLinertec
ModelLTS-205N
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals